DT5-6 low-speed centrifuge is a commonly used centrifuge in laboratory, hospital laboratory, emergency laboratory, township hospital, community hospital, etc. It is small in size, cheap in price and stable in quality.
The maximum speed is 4000 rpm, the rotor capacity is 10/15ml×8, and the commonly used vacuum blood collection tubes are applicable.
DT5-6 low speed centrifuge fault code
When the system is faulty, the Time window will display the corresponding fault code.
error code | Fault type | Solution |
E2 | Overspeed fault | Power on again |
E3 | Door cover is not closed | Close the door and start again |
E4 | Running door open | Re-close the door |
E7 | Zero speed failure | Check if the motor connector is loose and the motor is blocked. |
Ed | Parameter not confirmed, press "Start" | After confirming the parameters, press the start again (automatically confirm the parameters after 3 seconds without the button operation) |
EE | Overload fault | Restart the machine |
EF | Motor stall | Rotate the rotor a bit; remove foreign matter from the blockage. |
EH | Hall failure | Check if the motor interface is loose |