DT5-6 low speed centrifuge failure description

DT5-6 low-speed centrifuge is a commonly used centrifuge in laboratory, hospital laboratory, emergency laboratory, township hospital, community hospital, etc. It is small in size, cheap in price and stable in quality.
The maximum speed is 4000 rpm, the rotor capacity is 10/15ml×8, and the commonly used vacuum blood collection tubes are applicable.

DT5-6 low speed centrifuge fault code

When the system is faulty, the Time window will display the corresponding fault code.

error code

Fault type

Solution

E2

Overspeed fault

Power on again

E3

Door cover is not closed

Close the door and start again

E4

Running door open

Re-close the door

E7

Zero speed failure

Check if the motor connector is loose and the motor is blocked.

Ed

Parameter not confirmed, press "Start"

After confirming the parameters, press the start again (automatically confirm the parameters after 3 seconds without the button operation)

EE

Overload fault

Restart the machine

EF

Motor stall

Rotate the rotor a bit; remove foreign matter from the blockage.

EH

Hall failure

Check if the motor interface is loose