1.    Components
The main components are shown in the following table and figure.
Main components |
Name of components |
Specification |
Quantity |
Remark |
DR system |
Portable high frequency x-ray machine |
MCX-P40-R |
1 set |
 |
Digital detector |
Xmaru1210p |
1 set |
 |
|
Synchronous box |
YJT-III |
1 set |
 |
|
Laptop |
Y470N-IFIÂ (H) |
1 set |
 |
|
DR image processing software |
-- |
1 set |
 |
|
Accessory equipment |
Machine frame |
YJF-I |
1 set |
Optional components |
Â
2.    Application
4kW digital portable medical diagnosis high frequency X-ray machine, used for human limbs checking and diagnosing, especially in the wild places, battlefield flow, playground, pet clinic, and rescue diagnosis etc.
Â
3.    Characteristics
1)Â Â Â Â Deft and concise.
2)Â Â Â Â 6 periods of digital LED display.
3)Â Â Â Â 24 kinds of presupposed anatomy memory pattern choices.
4)Â Â Â Â Accuracy control method of simulation and double digital loops.
5)Â Â Â Â High precision control of tube voltage, tube current.
6)Â Â Â Â Failure self-protection, self-diagnosis.
4.    Dimension
(1)  X-ray machine head (length × width × height): 442mm × 347mm × 230mm
(2)  Machine stand (length × width × height): 1060 mm × 790 mm × 2210 mm
5.    Parameters
 Model |
YJP040DR-A |
Input power |
Voltage: AC220V±22V  Frequency: 50/60Hz±1Hz      Power capacity: ≥5kVA   Inner resistance: ≤0.5Ω |
Maximum output power |
P =110kV×36mA=4kW |
Nominal power |
100kV, 40mA, 0.1s, 4kW |
Frequency |
≥40kHz |
kV adjusting range |
40kV~110kV, continuous adjustment step 1Â kV |
mA adjusting range |
36mA~70mA |
mAs adjusting range |
1mAs-200mAs, shifted adjustment |
s adjusting range |
0.04s~3.2s |
X-ray tube |
Focus: 0.6/1.5mm;Â Anode heat capacity:Â 76kHU |
Varian digital detector |
Limit resolution:Â 3.6LP/mm;Â A/D conversion:14bit; Pixel area:427*356mm;Â Pixel matrix:Â 3072*2560 |
Laptop |
a) Model:Â Lenovo Y470N-IFI(H)Â Â Â Â b) CPU:Â Intel Core i5-2410M(2.3GHz) c) Memory capacity:Â 4GBÂ d)Â Hard drive capacity:Â 750GBÂ e) Screen size: 14 inch |
Anatomy program |
48 kinds of memory choices |
X-ray machine Size |
290×260×230 mm |
Weight |
18.8kg |
Accessory |
Standard: switch, cables; Option: mobile scaffold |
Machine frame |
The highest size of mobile frame is: 1060 × 790 × 2210mm The lowest size of mobile frame is: 1060×790×1240mm The net weight of mobile frame about: 23 kg |
Infusion and injection products are medical devices that are administered through either infusion or injection methods to deliver therapeutic agents into the body. These products are commonly used in hospitals, clinics, and other healthcare settings for various purposes, such as hydration, medication administration, blood transfusions, and nutrition support, such as I.V cannula, infusion set, syringe, etc.
Infusion set,injection products,iv set,iv infusion set,i.v cannula
2 MEDS TECHONOLOGY CO.,LTD , https://www.mtclosedsuction.com